Validating the InstrumentHub
This page is for QA staff who validate the InstrumentHub for GxP use. It describes who tests what, what Scifeon’s own tests of the InstrumentHub cover and where they stop, and suggested steps to qualify an InstrumentHub in your laboratory. For how the InstrumentHub is connected and secured, see IT and security.
Who tests what
Section titled “Who tests what”| Part | Covered by |
|---|---|
| The File Hub software, in the file gateway role | Scifeon’s automated supplier tests of the File Hub, and the Scifeon platform validation package for the Scifeon side of file ingestion. See Validating the File Hub. |
| The instrument reader software | Scifeon’s automated tests, run for every change to it. See What the supplier tests cover. |
| The preparation and updates of the InstrumentHub | Scifeon’s automated tests of its tools, and a check that each InstrumentHub runs when its security baseline is applied. |
| Each instrument model connected to the reader | Connected by Scifeon or by you when the model is first used. The instrument pages list the cable and settings for the models connected so far. |
| The InstrumentHub in your laboratory, with your instruments, instrument PC and settings | Your installation and operational qualification (IQ/OQ). Qualifying an InstrumentHub has suggested steps. |
What the supplier tests cover
Section titled “What the supplier tests cover”Instrument reader
Section titled “Instrument reader”Automated tests run for every change to the instrument reader. They show that the value and unit are taken from the instrument’s output with the configured patterns, also over several lines, and that a reading that does not complete in time is kept but marked as incomplete.
Preparation and updates
Section titled “Preparation and updates”Scifeon’s tools to prepare and update InstrumentHubs are tested, and each time the security baseline is applied the InstrumentHub checks it; in an update, an InstrumentHub that fails the check is reported to Scifeon and not restarted.
Limits of the supplier tests
Section titled “Limits of the supplier tests”The instrument reader is tested with sample output, not with a connected instrument. Each instrument model is checked when Scifeon or a customer first connects it, and you verify your own instruments in your OQ.
A detailed description of the test coverage and its limits is part of the supplier documentation described at the top of this page.
Qualifying an InstrumentHub
Section titled “Qualifying an InstrumentHub”The steps in this section are a suggestion from Scifeon. You decide how to qualify the InstrumentHub, based on your own validation approach and risk assessment, and you can change, add or leave out steps.
Record the result, the date and the person for each step.
Installation qualification (IQ)
Section titled “Installation qualification (IQ)”| # | Step | Expected result |
|---|---|---|
| IQ1 | Check the name on the InstrumentHub against the name Scifeon gave you. | They match. |
| IQ2 | In Scifeon, open the InstrumentHub’s hub. | The hub is listed under the InstrumentHub’s name. |
| IQ3 | File gateway: connect the instrument PC to the InstrumentHub’s network port with a network cable. The InstrumentHub gives the PC its address, see The network link. Run ipconfig /all on the PC. | That adapter has an address from the InstrumentHub, between 172.31.250.50 and 172.31.250.150, or the fixed address you set in 172.31.250.0/24. It has no default gateway and no DNS server from the InstrumentHub. |
| IQ4 | File gateway: on the instrument PC, run Test-NetConnection 172.31.250.1 -Port 445 and Test-NetConnection 172.31.250.1 -Port 22 in PowerShell. | Port 445 (the share) succeeds. Port 22 (SSH) fails. |
| IQ5 | File gateway: open \\hub012\drop on the PC without a user name, then with the share’s user name and password from Scifeon. | Without credentials, access is refused. With them, the share opens. |
| IQ6 | File gateway: open http://hub012.local/ on the PC and add the environment, unless Scifeon has done so. | The environment shows Connected. |
| IQ7 | List who can sign in to the instrument PC. | Recorded. |
Operational qualification (OQ)
Section titled “Operational qualification (OQ)”| # | Step | Expected result |
|---|---|---|
| OQ1 | Instrument reader: connect the instrument as described on its instrument page and configure the device in Scifeon. | The instrument appears in Scifeon as a device of the hub. |
| OQ2 | Instrument reader: take a reading on the instrument. | The value and unit shown in Scifeon equal those shown on the instrument. |
| OQ3 | Instrument reader: repeat OQ2 for each kind of reading you use, for example each unit or method. If the instrument allows it, also interrupt a reading. | As OQ2. An interrupted reading is shown in Scifeon as incomplete. |
| OQ4 | File gateway: run the File Hub OQ, with the watched folder on the drop share. | As in the File Hub OQ. |
| OQ5 | File gateway: set Folder shown to users on the watched folder’s device in Scifeon, see File gateway. | Files are shown in Scifeon with that folder. |
| OQ6 | File gateway: ask Scifeon to turn off the InstrumentHub’s 4G connection for the test. Export a result, then have the connection turned back on. | The file waits on the InstrumentHub while the connection is off, and arrives in Scifeon once, with the same SHA-256, after it is back. |